Home / Chubb / Determination 806484
Determination 806484
Claim declined for damage to high voltage testers allegedly caused by power surge
Summary
A company insured under an equipment policy claimed for damage to two high voltage testers, alleged to be from a power surge, after the insurer declined the claim. AFCA found the complainant had not established an identifiable single incident and that the damage was likely caused by excluded internal breakdown or malfunction, so determined in favour of the insurer.
The policy covers accidental damage from a single identifiable incident and excludes breakdown, malfunction or electrical supply other than a power surge, and the complainant did not establish a power surge.
Issues
- Claim denial primary · not upheld
“The insurer is not required to take any further action in relation to this claim; claim decline upheld as complainant failed to establish identifiable single incident.”
- Inadequate claim investigation secondary · not upheld
“The available information shows the insurer met its obligations to fairly assess the claim.”
Subjects and remedies
business interruptionno remedy
Text of the determination
Determination
Case number 806484
Financial firm Chubb Insurance Australia Limited
Determination | Case number: 806484 Page 1 of 5
Case number: 806484 25 February 2022
1 Determination overview
1.1 Complaint
The complainant holds an equipment insurance policy with the financial firm (insurer). The
policy lists a high voltage tester with a sum insured of $65,000. The policy also covers
unspecified equipment up to $3,000 per item.
The complainant lodged a claim for damage to two high voltage testing units, the listed high
voltage tester (voltage tester 1) and an unlisted voltage tester (voltage tester 2). The
complainant says on 7 February 2020 both units were damaged by a power surge.
The insurer declined the claim. It says:
the complainant has not established the damage was caused by a power surge, and
the material provided suggests the damage was caused by an internal breakdown or
malfunction which is excluded from cover.
The complainant is the insured company represented by a broker. Reference in this
determination to the complainant includes reference to the representative.
1.2 Issues and key findings
Is the insurer entitled to decline the claim?
Yes. The complainant has the onus to show a single identifiable incident caused accidental
damage to the equipment. The complainant has not established this. The exchanged material
shows it is more likely than not the damage was caused by breakdown, malfunction or
electrical supply, other than a power surge, during the equipment’s gen eral use. This is not
covered by the policy terms.
Why is the outcome fair?
The terms of the policy set out the parties’ responsibilities and obligations under the contract
of insurance. The available information shows the insurer met its obligations to fairly assess
the claim. It would not be fair to require the insurer to cover the claimed damage when there
is insufficient information to show it was caused by an incident covered by the policy.
1.3 Determination
This determination is in favour of the insurer.
The insurer is not required to take any further action in relation to this claim.
Determination | Case number: 806484 Page 2 of 5
2 Reasons for determination
2.1 Is the insurer entitled to decline the claim?
Yes. The complainant has the onus to show a single identifiable incident caused accidental
damage to the equipment. The complainant has not established this. The exchanged material
shows it is more likely than not the damage was caused by breakdown, malfunction or
electrical supply, other than a power surge, during the equipment’s general use. This is not
covered by the policy terms.
Complainant has initial onus to establish loss or damage covered by the policy
When lodging an insurance claim, the complainant has the initial onus of proof to establish,
on the balance of probabilities, they suffered loss or damage covered by the policy terms.
This is known as establishing a prima facie claim, or a claimable loss.
If the complainant can establish a claimable loss, the onus passes to the insurer to establish
an exclusion applies or the claim is otherwise not covered.
Policy covers accidental damage arising from an identifiable single incident
The complainant’s p olicy provides cover for accidental damage. The policy states:
[The insurer] will compensate [the complainant] on the terms and conditions of
this policy, for accidental damage to the equipment occurring by physical means.
The policy defines ‘accidental damage’ as:
physical damage which occurs as a result of a sudden, unforeseen and
unexpected event. The event must arise from a single identifiable incident.
Cover is, however, subject to any applicable policy exclusion. Relevant to this complaint, the
policy excludes cover for accidental damage to the equipment caused by ‘ servicing,
breakdown, malfunction, design fault or electrical supply other than a power surge’.
Complainant says it is likely a power surge caused the damage
The complainant initially lodged the claim for damage to voltage tester 1. The claim
lodgement form says there was an internal fault that trips the power supply when turned on.
The complainant first thought it might be due to recent transit or the age of the unit.
The complainant has since claimed for the unlisted voltage tester 2 as well. The complainant
says they did not know unlisted equipment was also covered. The complainant says after
voltage tester 1 failed they powered up voltage tester 2 however the V3 output did not work.
The complainant says it is likely a power surge occurred and caused damage to both units.
Complainant provided reports from the repairer
The complainant provided reports from the repairer (ET). ET says both electrical units had to
be sent to its factory in Italy for testing. For voltage tester 1, ET says its factory confirmed a
Determination | Case number: 806484 Page 3 of 5
number of boards were faulty (namely the V1-V2-V3 voltage amplifier codes and the internal
power supply module). ET says:
It is unclear as to what the cause of this sort of damage can be however from past
experience with repairs of this nature by a variety of brands and equipment that
we sell, it boils down to a number of possible causes.
Here are some possibilities:
Mains borne spikes
Static on test leads and cables,
Ground potential differences,
to name a few.
For voltage tester 2, ET says there was a V3 voltage generator overload when energised and
an internal inspection shows discolouration to the PC board. ET says:
The factory advised our Service Department the following:
On this unit they had to replace the measuring transformer of V4 and
repair the measuring circuit of 10 (Input/Output).
Both of faults could be due to some form of electrical stress applied to the
unit like fast transients, surges, floating grounding, etc.
ET again advised it is unclear what caused this type of damage, but it could be from any of
the possibilities already listed above.
ET also says the ‘spikes/conditions are hard to determine and at times even the most
seasoned practitioner can be unaware of their presence’.
The complainant says a power surge/spike is a sudden, unforeseen, and unexpected event
which is identified by ET as a realistic and strong probability for the cause of the damage. On
this basis the complainant says the claim should be covered.
Insurer’s assessor says the damage was caused by an internal fault
The insurer’s loss adjuster (JA) reviewed the available information and rep orted on the
claimed damage. While JA did not assess the damaged units, JA says the voltage testers are
specifically designed to accept high current and voltage which generates significant heat
during their usual use. JA says these items would not be considered susceptible to power
interruption and if a power surge had occurred you would expect damage to other, more
sensitive, electrical equipment. There is no evidence of this.
JA says it requested further information from the complainant, such as a letter from the power
company, to show a power interruption occurred. This has not been provided. JA concluded,
given the lack of evidence to show a power surge occurred, the more probable cause of
damage is an inherent internal failure of the electrical components within each tester.
The insurer says the complainant has not established a claimable loss but even if they had,
JA’s assessment supports a conclusion that the damage was caused by an internal
breakdown, malfunction or fault and the claim is therefore excluded.
Determination | Case number: 806484 Page 4 of 5
Insurer entitled to decline the claim
There is no dispute there is damage to both voltage testing units. I accept this damage was
sudden, unforeseen and unexpected for the complainant. However, this is not enough to
establish a claimable loss. The policy terms require there to be an identifiable incident that, on
balance, caused the accidental damage. This requires more than a list of potential causes.
ET does not identify any single incident as the more probable cause, but instead lists several
possibilities. ET clarified the suggested list was ‘to name a few’ which means the list is not
exhaustive.
I am not satisfied the exchanged information establishes it is more likely than not that any
single identifiable incident caused the claimed damage.
Even if I accept ET’s reports establish accidental damage caused by a single identifiable
incident (which I do not), the available cover is subject to the policy exclusions. The policy
excludes cover for damage caused by breakdown, malfunction and ‘ electrical supply other
than a power surge ’ . ET’s list of possible causes includes other electrical supply issues which
do not directly relate to a power surge.
In the absence of sufficient information to show a power surge occurred, I accept the
information suggests there has been an internal malfunction or breakdown during the
equipment’s general use. This is not covered by the policy terms.
2.2 Why is the outcome fair?
The terms of the policy set out the parties’ responsibilities and obligations under the contract
of insurance. The available information shows the insurer met its obligations to fairly assess
the claim. It would not be fair to require the insurer to cover the claimed damage when there
is insufficient information to show it was caused by an incident covered by the policy.
Insurer fairly considered the claim
Insurance policies do not cover all circumstances that give rise to a loss. The insurer is
entitled to rely on the terms of the policy and the available information when assessing the
claim.
There is no information available to me to show the insurer made an error or incorrectly
assessed the claimed damage.
Determination | Case number: 806484 Page 5 of 5
3 Supporting information
3.1 The AFCA process
AFCA’s approach is based on fairness
AFCA has determined this complaint based on what is fair in all the circumstances, having
regard to:
the legal principles
applicable industry codes or guidance
good industry practice
previous decisions of AFCA or its predecessor schemes (which are not binding).
The respective parties have completed a full exchange of the relevant information, and each
party has had the opportunity to address any issues raised. We have reviewed and
considered all of the information the parties have provided.
While the parties have raised a number of issues in their submissions, we have restricted this
determination to the issues that are relevant to the outcome.
We assess complaints on available information and circumstances
AFCA is not a court of law. We do not have the power to take or test evidence on oath, or to
require third parties to give evidence.
When we assess complaints, we consider:
available documents
the recollections of the parties
all relevant circumstances.
We give more weight to documents created at the time the events occurred. If there are no
relevant documents, we will decide what most likely occurred based on the available
information.
If there are conflicting recollections and these are evenly weighted, we may find that a claim
cannot be established.