Determination 363715
Denial of claim
Summary
The applicant claimed for damage to an ultrasound machine under an asset protection policy. The FSP declined as the loss arose from an internal fault, which is excluded, and AFCA determined the decline was correct.
The policy covers only external causes, and the applicant did not establish on the balance of probabilities that the loss arose from an external event.
Issues
- Claim denial primary · not upheld“The FSP was entitled to decline the claim because the cause of the loss is specifically excluded by the terms of the policy.”
- Inadequate claim investigation secondary · not upheld“That he did not personally inspect the machine does not prevent him from assessing the likely cause of the failure”
Subjects and remedies
no remedy
Related
All Financial Firm (Unknown - Insurer) determinations · Find similar determinations
Text of the determination
Determination
General Insurance - Small Business/Farm Insurance - Commercial Property - FSP
Decision - Denial of claim-Exclusion/ condition
Determination
Case No: 363715 Determination Page 1 of 7
Case number : 363715 9 January 2015
1 Overview
1.1 Dispute
The Applicant is a beneficiary of an asset protection insurance policy underwritten
by the Financial Services Provider (FSP). The po licy covers a Phillips IE33
ultrasound machine.
On or about 5 December 2013 the FSP was notified of a claim for damage to the
machine.
The FSP denied the Applicant’s claim on the basis that the loss was excluded by
the terms of the policy.
1.2 Issue and key findings
Was the FSP entitled to decline the claim?
The FSP was entitled to decline the claim because the cause of the loss is
specifically excluded by the terms of the policy.
1.3 Determination
This Determination is in favour of the FSP.
The FSP was entitled to decline the claim. No further action is required on behalf
of the FSP.
Case No: 363715 Determination Page 2 of 7
2 Reasons for Determination
2.1 Was the FSP entitled to decline the claim?
The Applicant is required to establish a prima facie claim
The policy provides cover for accidental loss or damage to the machine from
external causes, subject to any exclusions.
It is accepted Insurance law that the Applicant carries the initial onus of proving
that an event covered under the terms of the policy has occurred. This is known
as a prima facie claim.
Once the Applicant has established a prima facie claim, the onus shifts to the FSP
if it wishes to deny the claim on the basis of an exclusion under the policy or some
other basis.
Expert evidence
The machine was inspected by the original equipment manufacturer (OEM) who
found no direct cause of failure. The fault was noted as being present in the upper
memory blocks (UMB), with no video signal evident at the point of failure.
However, the power supply unit (PSU) was still functional, indicating power was
still being provided to the machine.
The FSP appointed loss adjuster “JA” to review the information provided and
tender a technical opinion as to the cause of the loss.
JA considered it unlikely that a power surge had caused the damage to the
machine. This was on the basis that had a power surge occurred, the PSU would
have been the first damaged component, as it was the initial connector point to
the mains supply.
That the PSU was still operational, suggests no power surge had occurred.
Rather, the fault appeared to lie with the UMB. Whilst the OEM recommended
replacement of the PSU, this was on a precautionary basis only.
The electricity supplier has confirmed there is no record of any power surges in
the area on the date in question. There is no evidence of any storm activity in the
area on the date of loss. Further, had the premises experienced a power surge, it
is likely that any other appliances plugged in at the time would also have been
affected. That there were reportedly no other items damaged, is suggestive of no
power surge occurring.
The Applicant has not provided any contrary expert evidence
The Applicant disputes JA’s findings and raises issue with the fact that JA did not
personally inspect the machine, but rather relied on the reports submitted by the
OEM.
The Applicant was given the opportunity by this Service to obtain his own report
from a suitably qualified person. However, the Applicant did not provide any
Case No: 363715 Determination Page 3 of 7
additional information in support of his case. As a result, JA ’s report is
unchallenged.
It is accepted that JA is appropriately experienced and qualified to comment in
this matter.
That he did not personally inspect the machine does not prevent him from
assessing the likely cause of the failure, having had recourse to review the
findings of the OEM.
In the absence of conflicting information, the FSP is entitled to rely on JA’s
findings that, on balance, the loss arose as a result of an internal fault with the
machine.
Analysis report
A copy of an analysis report undertaken over the period of 28 February 2013 to 4
March 2013, has been supplied. It is unclear who undertook this analysis although
it is assumed to be the OEM. It is not clear if JA was made aware of these reports.
The report shows that the system shutdown abnormally and was restarting
without having been shut down properly. The analysis shows three faults on 28
February 2013, one fault on 1 March 2013 and three faults on 4 March 2013.
This is suggestive of an ongoing internal issue with the machine.
The analysis details a number of potential causes and/or troubleshooting
techniques to remedy the problem.
It is of note that none of these items listed in the analysis were found to be evident
upon inspection by the OEM in October. The applicant has also not addressed the
issue of whether the unit remained operational between March and when the
OEM inspected it in October. On balance, it appears likely that it did remain
operational in the meantime.
It is unlikely that any external factor would cause the intermittent problems in
March and then lie dormant until October.
The policy covers loss arising from an external cause
The policy provides damage cover which is caused by external means. Therefore,
the damage must arise from some factor external to the machine itself. It is of
note that the precise cause of loss has been unable to be determined.
However, the evidence presented suggests that the loss arose as a result of the
failure inherent to the machine itself. That is, an internal cause. This is excluded
by the terms of the policy.
No evidence has been presented to suggest the loss arose from any other
external cause, for example, impact to the machine or malicious damage. Whilst
the OEM report does not conclusively establish the cause of the loss, it is of note
the purposes of the OEM’s report was not to establish whether an internal or
external fault caused the failure. Rather, their commission was to establish where
the fault lay and how to remedy it.
Case No: 363715 Determination Page 4 of 7
The unit was purchased in 2009 as an ex-demonstrator model. JA confirms that
the low-voltage reading from the PSU is most likely due to the drying out of the
electronic capacitors, which occurs due to prolonged use of the unit and is
considered to be general wear and tear experienced over its lifespan.
In any event, whilst the cause of the UMB failure is not been determined, there is
nothing to suggest that this arose as a result of any external factor.
After considering all of the available information, on balance, the loss is
attributable to an inherent failure within the machine itself, being an internal
cause. As such, the Applicant has not met the onus upon him to establish that the
loss arose from an event covered by the terms of the policy. That is, the Applicant
has not demonstrated that the loss arose from external causes .
On this basis, the FSP has correctly declined the claim.
Case No: 363715 Determination Page 5 of 7
3 Supporting information
3.1 FOS approach to this dispute
This dispute has not been the subject of a FOS Recommendation. Due to its
nature, it was appropriate to proceed directly to a Determination in line with the
Operational Guidelines to the FOS Terms of Reference.
There has been a full exchange of information between the parties and each part y
has had the opportunity to address any issues raised by the other. Material
supplied by the parties will only be referred to in this decision where relevant and
in context.
3.2 Expert Evidence
Record Key points
FSP’s expert ( J A)
Loss Adjuster’s report
27 Jun 2014
instructions were to undertake a technical review in
relation to potential causation and the appropriate policy
response
examination of the claim documents has been completed
we understand the unit was left on overnight as normal
practice to allow transfer of data
the following morning the unit would not restart and the
manufacturer (OEM) was requested to investigate
the OEM found a video signal to the monitor although the
power supply unit (PSU) was operating (albeit at a lower
level)
the upper me mory blocks (UMB) were not logging data
from the input system
the OEM attempted to repair the unit by replacing the
video drivers and upgrading to match the current
component firmware
this was unsuccessful in testing, revealing that the
complete UMB assemb ly would require replacement
along with the PSU on a precautionary basis
we understand that repairs have not yet been affected,
pending the outcome of this claim
the OEM advised in their report that there is no evidence
of a direct cause of breakdown
the i nsured has speculated that this occurred as a result
of a power surge on the night
records from the bureau of meteorology were obtained
which confirm that no storms or lightning were recorded
on or around the date of loss
there also appears to be an absenc e of reported damage
to any other equipment in the facility
when examining power fluctuations as the potential
damage trigger, it is logical to commence investigations
with the PSU
Case No: 363715 Determination Page 6 of 7
as the PSU is the first exposure to incoming mains
electricity supply, any adverse power event will affect this
component first
the PSU was reported a s not operating correctly,
operating with voltage lower than normal acceptable
limits
this type of fault is not consistent with exposure to
adverse voltage but rather indicative of internal
deterioration of the PSU components
the PSU contains electrolytic capacitors that will dry out
and lose efficiency after an extended period in service
particularly where the unit is powered 24 hours a day
as this process develops the tolerance be gin to drift and
the voltage output on the PSU will gradually fall below
normal operating limits
the PSU output was only marginally below the tolerance
limit for this unit and not sufficient to prevent correct
operation
had there not been the fault with UM B assembly, there is
little doubt that this machine would have operated
correctly
it is likely the PSU was in this condition prior to the
incident
whilst the cause of damage to the UMB assembly has not
been determined, the evidence indicates it is highly
u nlikely this is as a result of a power surge
had a power surge occurred, the adverse voltage would
need to pass through the PSU and upstream to UMB
without damaging other components
this is not a plausible occurrence particularly in the
absence of other eq uipment being affected
we consider the cause of the damage the UMB is as a
result of an internal electronic breakdown within the
machine operating components
Applicant’s expert
OEM service report
21 Oct 2013
system shutting down with no indication of direct cause
monitor displays no video signal at point of failure
PSU remains but UMB not logging
suspect UMB/video card failure
try to repair using new 9800 video cards including
software system upgrade for compatibility
if problem persists, new UMB assembly unit will be
required
Applicant’s expert
OEM Service Report
24 Oct 2013
original problem reported was system shutting down with
no indication of cause
monitor displays no video signal at point of failure
PSU remains but UMB not logging
new vi deo cards installed and reloaded new software
system but system still failed to process correctly
new UMB assembly required
also marginal supply below system operating
Case No: 363715 Determination Page 7 of 7
specifications, replacement of PSU advised
Applicant’s expert
Analysis Report
29 Apr 2013
the system shutdown abnormally
the system is starting up without having been shut down
properly
corrective action includes:
asked if the site had lost power or if the user turn the
power of for any reason
various issues can cause a system to hang requiring the
user to switch off the power
verify that no one entered Tech admin in the setups,
which can reset event logs and incorrectly indicate an
abnormal shutdown
assuming that the system powered off by itself , likely
cause is a defective PSU or a de fective PCB
with such failures the system loses all power and causes
an indicator to flash
depending on the cause there may or may not be a user
recovery
a hard fault is not recoverable
a marginal voltage or overcurrent error may allow a
circuit breaker cy cle to reset supply and allow the system
to restart
majority of overcurrent faults are due to either a defective
channel board or SPD assembly
the majority of power supply voltage faults are due to
variation of the voltage which may be corrected by
readjus ting the supply
replace the defective component or perform power
supply adjustment